KOBRA® CELL Membrane
Art. No. RBRK02
Product info
Intended use
Replacement membrane for the KOBRA® CELL (RBRK01).
General Information
It is necessary to regularly monitor the performance of the KOBRA® CELL in order to detect any deterioration in the membrane contained within. The performance should be checked at the time of installation and then weekly by comparing the peak areas of a known aflatoxin standard. The same should also be done in order to monitor the deterioration of the lamp in the detector. The membrane is the site of the actual derivatisation reaction (ion-exchange occurs on the membrane). The membrane starts a clear colour and changes to yellow in colour around the reaction centre. The colour change indicates that the membrane will need to be replaced. Deterioration of the membrane will occur over a period of time depending upon the frequency of use and the type of samples analysed. When the performance becomes unacceptable the membrane should be replaced. Normally, it is found that even under extreme workloads the membrane does not need to be replaced for at least 6 months or 1,000 injections.
.
Accessories
Picture galleryDear customers,
we now provide the documents for our products in an electronic format which include the Instructions for Use (IFU), Safety Data Sheets (SDS), Application notes and the Certificate of Analysis (CoA). For batches placed on the market after 1st June 2023, you can find our documents on the eIFU portal eifu.r-biopharm.com/rbr.
Specifications | |
---|---|
Art. No | RBRK02 |
Test format | 1 x replacement membrane |