SureFood® ALLERGEN Peanut
Art. No. S3603
Product info
Intended use
The real-time PCR test detects DNA of peanut (Arachis hypogaea) according to directive (EC) 1169/2011 qualitatively and / or quantitatively. Each reaction contains an internal amplification control (IAC).
For the quantitative determination the use of the laboratory reference material SureFood® QUANTARD Allergen 40 (Art. No. S3301) containing 40 mg peanut / kg food sample and a detection system including a standard series is required.
General Information
The real-time PCR assay can be performed with commonly used real-time PCR instruments, equipped for detection of two fluorescence emissions at the channels FAM and VIC/HEX at the same time. The technical verification of instruments was performed on Roche LightCycler® 2.0, Roche LightCycler® 480 II, Roche cobas® z 480 Analyzer, Qiagen Rotor-Gene Q, Applied Biosystems 7500, Bio-Rad CFX96, Bio Molecular Systems MIC, Agilent AriaDx and Agilent Mx3005P.
Accessories
Dear customers,
we now provide the documents for our products in an electronic format which include the Instructions for Use (IFU), Safety Data Sheets (SDS), Application notes and the Certificate of Analysis (CoA). For batches placed on the market after 1st of January 2024, you can find our documents on the eIFU portal congen.de/information.
Specifications | |
---|---|
Art. No | S3603 |
Test format | 100 reactions |
Sample preparation | For DNA-preparation the use of SureFood® PREP Advanced (Art. No. S1053), protocol 1 or SureFast® Mag PREP Food (Art. No. F1060) is recommended. |
LOD (Detection Limit) | The SureFood® ALLERGEN Peanut PCR has a limit of detection of ≤ 0.4 mg / kg using SureFood® PREP Advanced, protocol 1. |
LOQ (Limit of quantification | 1 mg / kg using SureFood® PREP Advanced, protocol 1 |