Intended use:

The real-time PCR test detects soya DNA (Glycine max) according to directive (EC) 1169/2011 qualitatively and / or quantitatively. Each reaction contains an internal amplification control (IAC).

General information:

For the quantitative determination the use of the laboratory reference material SureFood® QUANTARD Allergen 40 (S3301) containing 40 mg soya / kg food sample and a detection system including a standard dilution is required.

For DNA preparation, the use of SureFood® PREP Advanced (S1053), protocol 1, is recommended.

The real-time PCR assay can be used with established real-time PCR instruments capable of detection of at least two fluorescence dyes. Internal validation was performed on Roche LightCycler® 480 II, Roche cobas® z 480 Analyzer, Qiagen Rotor-Gene Q, Applied Biosystems 7500, Bio-Rad CFX96, Bio Molecular Systems MIC, LTF MyGo Pro, AriaDx and Agilent Mx3005P.

The method is validated for quantitative determination of 1 mg to 400 mg allergenic substance / kg food sample using the reference material SureFood® QUANTARD Allergen 40 containing 40 mg allergenic substance / kg food sample.

Watch our video tutorials on sample preparation and test procedure with SureFood® ALLERGEN Soya:

.

Accessories:

Specifications
Art. No. S3601
Test format 100 reactions
LOD (Detection limit) ≤ 0.4 mg / kg
LOQ (Limit of quantification)

1 mg / kg using SureFood® PREP Advanced, protocol 1

Evaluation Real-time PCR instrument with two detection channels (510 nm and 580 nm)
E-mail linksales@r-biopharm.de
Files
InstructionsGerman/English
SDSerklaerung-zu-congen-msds_2018-12.pdf (German)
confirmation-congen-msds_2018-12.pdf (English)
Similar products
RIDASCREEN®FAST Soya
RIDASCREEN®FAST Soya
RIDASCREEN®FAST Soya is a sandwich enzyme immunoassay for the quantitative analysis of native and processed soya proteins in food. Such products ar...
RIDA®QUICK Soya
RIDA®QUICK Soya
RIDA®QUICK Soya is an immunochromatographic test for the qualitative detection of soya contamination (native and processed) on surfaces (swab test ...
Set of 3 processed Soya Assay Controls
Set of 3 processed Soya Ass...
The processed assay controls are used to control extraction, test performance and handling of the RIDASCREEN®FAST Soya (R7102) for detection of Soya.

Product support

Questions? Tap into our team’s expertise. We’re here to support you and your business throughout the testing process to ensure your success.

Start typing and press Enter to search